How often to change Barnstead FL1192x1 Ultrafiter Membrane.
Date : 9/14/2020
From :
eric
Question
how often am i supposed to change the FL1192x1 filter cant find any details on that filter
Answer
The UF membrane is very durable and can last over 5 years depending on what is in the water you are feeding to it. Your best bet is to do an LAL test looking for pyrogens, If you see them after sanitization, they membrane most likely needs to be changed. Short of that, I would replace it at least every 3 years if you are doing work that is effected by endotoxins.
This high capacity four housing lab water system is designed to provide up to 3 l.p.m. of ultrapure water exceeding ASTM and CAP Type I, II or III specifications.
Code F, 2.5 inch X 10 inch. Coconut shell carbon block with 10 micron outer wrap. Special Prep 1, for Chlorine removal with low extractable applications - 1/pk
Filter Housing
IWT 3C0600002 Style High Capacity Filter - Cartridge Universal II
This High Capacity Mixed Bed Deionization Filter is used for the reduction of dissolved minerals. with Color Indicator - Comparable to IWT and US FILTER 3C0600002 and HC200 or HC-200
Great for Lab Water Pretreatment
or Great Drinking Water
Images are representative of the products. Images may or may not be of the actual product. If it is important e-mail us for an actual image if available.
* Flat Rate UPS shipping when able to ship via UPS and is in the USA excluding Hawaii and Alaska. Larger Items may not be able to ship via UPS, in that case freight charges will be quoted seperately. International shipping will be quoted after the order is placed. You will have the opportunity to cancel before we finalize your order.